用于旋转头部运动测量的可穿戴仪器头带的实验室评估
Laboratory evaluation of a wearable instrumented headband for rotational head kinematics measurement
摘要 Abstract
轻度创伤性脑损伤(mTBI)是一种高发性疾病,不同个体的预后具有异质性。影响脑组织变形及mTBI风险的关键因素之一是头部的旋转运动学特性。由于传感器与颅骨的良好耦合,仪器化牙套被广泛接受为测量旋转头部运动的方法。然而,在某些情况下,例如长期数据收集时,佩戴牙套并不总是可行的。因此,需要替代的可穿戴设备。本研究提出了一种改进设计和数据处理方案,用于一种仪器化头带。我们的仪器化头带利用一组惯性测量单元(IMU),并采用基于连续小波变换的新数据处理方案,以解决IMU测量中的误差来源。在人体模型试验装置上的实验室测试表明,当头带受到足球撞击以模拟头球动作时,与参考传感器相比,头带测得的前部撞击峰值旋转速度(PRV)和峰值旋转加速度(PRA)的相关系数(r)分别为0.80和0.63,归一化均方根误差(NRMSE)分别为0.20和0.28。然而,考虑所有撞击位置时,相关系数r分别下降至0.42和0.34,NRMSE分别增加至0.5和0.41。这种新的仪器化头带在重建由正面足球撞击引起的旋转头部运动学方面优于先前的设计,为仪器化牙套提供了一种潜在的替代方案。
Mild traumatic brain injuries (mTBI) are a highly prevalent condition with heterogeneous outcomes between individuals. A key factor governing brain tissue deformation and the risk of mTBI is the rotational kinematics of the head. Instrumented mouthguards are a widely accepted method for measuring rotational head motions, owing to their robust sensor-skull coupling. However, wearing mouthguards is not feasible in all situations, especially for long-term data collection. Therefore, alternative wearable devices are needed. In this study, we present an improved design and data processing scheme for an instrumented headband. Our instrumented headband utilizes an array of inertial measurement units (IMUs) and a new data-processing scheme based on continuous wavelet transforms to address sources of error in the IMU measurements. The headband performance was evaluated in the laboratory on an anthropomorphic test device, which was impacted with a soccer ball to replicate soccer heading. When comparing the measured peak rotational velocities (PRV) and peak rotational accelerations (PRA) between the reference sensors and the headband for impacts to the front of the head, the correlation coefficients (r) were 0.80 and 0.63, and the normalized root mean square error (NRMSE) values were 0.20 and 0.28, respectively. However, when considering all impact locations, r dropped to 0.42 and 0.34 and NRMSE increased to 0.5 and 0.41 for PRV and PRA, respectively. This new instrumented headband improves upon previous headband designs in reconstructing the rotational head kinematics resulting from frontal soccer ball impacts, providing a potential alternative to instrumented mouthguards.